The aim of PhD dissertation was to identify disturbances which may occur in the calibration facility for instruments measuring the volumetric flow rate or mass flow rate and the quantity (volume or mass) of gas and to determine the effects of disturbances on the accuracy of the facility. In the PhD dissertation analysis of the possible location for the calibration facility in the gas compressor station, taking into account the occurrence of pulsations in the pressure and flow rate. To identify disturbances and to optimize the construction of the facility, computational fluid dynamics methods were used. The results of the numerical simulations were verified in a dedicated laboratory facility. The laboratory tests were performed using hot-wire anemometer and turbine meters. The results of this work were used to design innovative on a world scale and the first high pressure calibration facility for gas meters in Poland which are used for billing purposes in natural gas transmission and distribution systems.
Record ID
WUTac395bbc03864c9fa41f319799f34e70
Diploma type
Doctor of Philosophy
Author
Paweł Szufleński Paweł Szufleński,, Faculty of Mechatronics (FM)
Title in Polish
Wpływ zaburzeń generowanych przez instalacje na dokładność wzorcowania przyrządów do pomiaru strumienia i ilości gazu
Language
(pl) Polish
Certifying Unit
Faculty of Mechatronics (FM)
Discipline
mechanical engineering / (technology domain) / (technological sciences)
Status
Finished
Start date
19-01-2011
Defense Date
12-03-2014
Title date
26-03-2014
Supervisor
Internal reviewers
External reviewers
Stanisław Waluś, prof. Stanisław Waluś, prof.,, External affiliation of publication: Silesian University of Technology
Honored
yes
Pages
148
Keywords in English
flow profiles, high pressure calibration facility, gas meter
Abstract in English
The aim of PhD dissertation was to identify disturbances which may occur in the calibration facility for instruments measuring the volumetric flow rate or mass flow rate and the quantity (volume or mass) of gas and to determine the effects of disturbances on the accuracy of the facility. In the PhD dissertation analysis of the possible location for the calibration facility in the gas compressor station, taking into account the occurrence of pulsations in the pressure and flow rate. To identify disturbances and to optimize the construction of the facility, computational fluid dynamics methods were used. The results of the numerical simulations were verified in a dedicated laboratory facility. The laboratory tests were performed using hot-wire anemometer and turbine meters. The results of this work were used to design innovative on a world scale and the first high pressure calibration facility for gas meters in Poland which are used for billing purposes in natural gas transmission and distribution systems.
Thesis file
  • File: 1
    Rozprawa doktorska_Paweł Szufleński.pdf
Request a WCAG compliant version
Citation count
2

Uniform Resource Identifier
https://repo.pw.edu.pl/info/phd/WUTac395bbc03864c9fa41f319799f34e70/
URN
urn:pw-repo:WUTac395bbc03864c9fa41f319799f34e70

Confirmation
Are you sure?
Report incorrect data on this page