Electronic calibration unit for DC-8 GHz vector-network-analyzer measurements

Michał Abramowicz , Arkadiusz Lewandowski

Abstract

e present design and realization of a vector- network-analyzer (VNA) electronic-calibration unit operating from DC to 8 GHz. Calibration standards are implemented on a printed circuit board and are connected to the VNA ports through a set of micro-electro-mechanical-system (MEMS) switches. We verify our unit by comparing corrected measure- ments of verification items obtained after calibrating the VNA with our electronic unit and with mechanical devices. The results show a very good agreement.
Author Michał Abramowicz ISE
Michał Abramowicz,,
- The Institute of Electronic Systems
, Arkadiusz Lewandowski ISE
Arkadiusz Lewandowski,,
- The Institute of Electronic Systems
Pages1-4
Publication size in sheets0.5
Book Organizing Committee of Conference MIKON: Proc. 21st International Conference on Microwaves, Radar and Wireless Communications, 2016, IEEE, ISBN 9781509022151, 732 p.
Keywords in Englishmicrowave measurements, vector network an- alyzer (VNA), calibration, electronic calibration unit, through- open-short-match (TOSM) calibration, thru-reflect-match (TRM) calibration
DOIDOI:10.1109/MIKON.2016.7491937
URL http://ieeexplore.ieee.org/xpl/login.jsp?tp=&arnumber=7491937&url=http%3A%2F%2Fieeexplore.ieee.org%2Fstamp%2Fstamp.jsp%3Ftp%3D%26arnumber%3D7491937
Languageen angielski
File
Lewandowski_A_mikon_2016.pdf (file archived - login or check accessibility on faculty) Lewandowski_A_mikon_2016.pdf 1.95 MB
Score (nominal)15
ScoreMinisterial score = 15.0, 27-03-2017, BookChapterMatConf
Ministerial score (2013-2016) = 15.0, 27-03-2017, BookChapterMatConf
Citation count*0
Cite
Share Share



* presented citation count is obtained through Internet information analysis and it is close to the number calculated by the Publish or Perish system.
Back