Microwave Determination of Liquid Mixing Ratio for Microfluidics

Paweł Barmuta , Bao Juncheng , T. Markovic , B. Nauwelaers , Dominique M. M.-P. Schreurs , I. Ocket

Abstract

This paper presents a microwave-based method for extracting liquid mixing ratio of two unknown liquids in microfluidics systems. The method is based purely on measurements, and no electromagnetic simulations nor liquid permittivity models are required. The method was successfully evaluated for three mixtures of isopropanol and water with varying concentration. The maximum error of extracting mixing ratio is 6.09%, which is probably mostly caused by fast isopropanol evaporation.
Author Paweł Barmuta (FEIT / PE)
Paweł Barmuta,,
- The Institute of Electronic Systems
, Bao Juncheng
Bao Juncheng,,
-
, T. Markovic
T. Markovic,,
-
, B. Nauwelaers
B. Nauwelaers,,
-
, Dominique M. M.-P. Schreurs
Dominique M. M.-P. Schreurs,,
-
, I. Ocket
I. Ocket,,
-
Pages745-746
Publication size in sheets0.3
Book Proc. MIKON 2018, 2018, Fundacja Mikrofal i Radiolokacji MIKON, ISBN 978-83-949421-0-6
Keywords in Englishmicrofluidics, liquid mixing, permittivity
DOIDOI:10.23919/MIKON.2018.8405280
URL https://ieeexplore.ieee.org/document/8405280/
Languageen angielski