Design and characterization of novel all-solid-state potentiometric sensor array dedicated to physiological measurements

Renata Toczyłowska-Mamińska , Monika Kloch , Anna Zawistowska-Deniziak , Agnieszka Bala

Abstract

A novel construction of all-solid-state potentiometric sensor array designed for physiological measurements has been presented. The planar construction and elimination of liquid phase creates broad opportunities for the modifications in the sensing part of the sensor. The designed construction is based on all-solid-state ion-selective electrodes integrated with the ionic-liquid based reference electrode. Work parameters of the sensor arrays were characterized. It has been shown that presented sensor design indicates high sensitivity (55.2±1 mV/dec, 56.3±2 mV/dec, 58.4±1 mV/dec and 53.5±1 mV/pH for sodium-, potassium-, chloride- and pH-selective electrodes, respectively in 10−5–10−1.5 M range of primary ions), low response time (t95 did not exceed 10 s), high potential stability (potential drift in 28-h measurement was ca. ±2 mV) and potential repeatability ca. ±1 mV. The system was successfully applied to the simultaneous determination of K+, Cl−, Na+ and pH in the model physiological solution and for the ion flux studies in human colon epithelium Caco-2 cell line as well.
Author Renata Toczyłowska-Mamińska
Renata Toczyłowska-Mamińska,,
-
, Monika Kloch
Monika Kloch,,
-
, Anna Zawistowska-Deniziak
Anna Zawistowska-Deniziak,,
-
, Agnieszka Bala ZMB
Agnieszka Bala,,
- Department Of Microbioanalytics
Journal seriesTalanta, ISSN 0039-9140
Issue year2016
Vol159
Pages7-13
Publication size in sheets0.5
Keywords in EnglishAll-solid-state potentiometric system; All-solid-state sensor array; Sensor array; Potentiometric sensor array; Ion transport; Ion flux
DOIDOI:10.1016/j.talanta.2016.06.001
URL http://www.sciencedirect.com/science/article/pii/S0039914016304180
Languageen angielski
File
Design and characterization of novel all-solid-state potentiometric sensor array dedicated to physiological measurements.pdf (file archived - login or check accessibility on faculty) Design and characterization of novel all-solid-state potentiometric sensor array dedicated to physiological measurements.pdf 909.99 KB
Score (nominal)40
ScoreMinisterial score = 40.0, 28-11-2017, ArticleFromJournal
Ministerial score (2013-2016) = 40.0, 28-11-2017, ArticleFromJournal
Publication indicators WoS Impact Factor: 2016 = 4.162 (2) - 2016=3.841 (5)
Citation count*0