Abstract

A multi-port reflectometer which allows us to determine a complex reflectance in pulse mode of the device under test is proposed. The reflectometer is based on a symmetrical five-port junction, directional coupler and diode detectors. The hardware configuration of the six-port junction and its calibration as well as the linearization of the diode detectors have been investigated. The experimental results for several terminations measured by the proposed reflectometer working under pulse excitation are compared with data measured by a Hewlett-Packard network analyzer at the selected frequencies
Author Ryszard Michnowski (FEIT / RE)
Ryszard Michnowski,,
- The Institute of Radioelectronics
, Marek Kukier (FEIT / RE)
Marek Kukier,,
- The Institute of Radioelectronics
, A. Turowski
A. Turowski,,
-
Pages547-550 vol.2
Book 13th International Conference on Microwaves, Radar and Wireless Communications. 2000. MIKON-2000, vol. 2, 2000
Keywords in Englishcalibration, Circuits, complex reflectance, diode detectors, directional coupler, directional couplers, Envelope detectors, microwave measurements, multiport networks, multi-port reflectometer, Phase detection, power measurement, pulse excitation, Pulse measurements, reflectivity, reflectometers, six-port reflectometer, symmetrical five-port junction, terminations, Testing
DOIDOI:10.1109/MIKON.2000.913992